Where the EU AI Act meets MDR and IVDR: one route in

Medical device software often includes AI. Diagnostic algorithms, treatment decision support, imaging analysis tools and adaptive monitoring systems all rely on machine learning or statistical inference. The question for manufacturers is whether that AI falls under the EU AI Act as well as MDR or IVDR, and if so, what changes in the conformity assessment process. The answer depends on the AI's role in the product, the classification of the device, and the structure of the conformity route. This matters because the obligations are different, the documentation is broader, and the timeline is tighter than many teams expect.
When AI in a medical device becomes high risk under the AI Act
The EU AI Act categorises certain AI systems as high risk. For products already covered by MDR or IVDR, that categorisation is narrow and conditional. Article 6(1) of the AI Act states that an AI system is high risk where it is a safety component of a product, or is itself a product, covered by the legislation listed in Annex I, and that product must undergo third party conformity assessment. MDR and IVDR are both listed in Annex I, Section A. So the first gate is whether your AI is the device or a safety component of it. A safety component is one whose failure or malfunction can compromise the safety or performance of the device. That definition comes from the product legislation, not the AI Act.
The second gate is whether the device must undergo conformity assessment by a notified body. Under MDR, Class IIa, IIb and III devices require notified body involvement. Class I devices, unless they have a measuring function or are sterile, do not. The same logic applies to IVDR. If your device is Class A under IVDR and does not require notified body assessment, the AI inside it is not high risk under Article 6(1) of the AI Act, even if that AI is critical to the device's function. The AI Act does not impose its high risk obligations in that case. This is a deliberately aligned threshold, designed to avoid creating parallel assessment routes for products that already pass through notified bodies.
It is easy to assume that all AI in medical devices is automatically high risk under the AI Act. That assumption is wrong. The Article 6(1) test is precise and conditional. Apply it carefully, and document the reasoning. If your AI does not meet both conditions, the AI Act's high risk requirements do not apply, though you remain responsible for meeting MDR or IVDR obligations for software, including lifecycle management, clinical evaluation and post market surveillance.
| Condition | What it means |
|---|---|
| AI is the device or a safety component | The AI system is the medical device itself, or its failure would compromise device safety or performance |
| Device requires notified body assessment | The MDR or IVDR classification mandates third party conformity assessment (Class IIa, IIb, III under MDR; Class B, C, D under IVDR) |
| Both conditions must be met | If either condition is not satisfied, the AI is not high risk under Article 6(1) and AI Act high risk obligations do not apply |
How the conformity assessment regimes combine
When an AI system in a medical device is high risk under Article 6(1), the EU AI Act does not create a separate conformity assessment. Article 43(3) specifies that for high risk AI systems covered by Annex I Section A legislation, including MDR and IVDR, the provider follows the conformity assessment procedure of that legislation. The AI Act's requirements for high risk systems, set out in Articles 8 to 15, are assessed as part of that procedure. This is the single assessment route. It means the notified body responsible for your MDR or IVDR conformity assessment also evaluates compliance with the AI Act's high risk obligations.
The practical consequence is that your technical documentation must now address both regimes. MDR Annex II and IVDR Annex II set out the technical documentation requirements for medical devices. When the AI Act applies, you must also meet the requirements in Annex IV of the AI Act, which specifies the information to be included in the technical documentation for high risk AI systems. Annex IV covers the AI system's design, development and performance, including risk management, data governance, transparency, human oversight, accuracy, robustness and cybersecurity. These are not bolt-on sections. They integrate into the device documentation, particularly the software lifecycle, risk management and clinical evaluation sections.
Article 17 of the AI Act requires providers of high risk AI systems to establish a quality management system. For MDR and IVDR products, this obligation is met by the existing quality management system required under those regulations. There is no need for a separate AI Act quality system. However, the system must now document how the AI Act's specific requirements are addressed, including how data quality is maintained, how transparency obligations are discharged, and how post market monitoring captures AI system performance. The notified body will assess this as part of the MDR or IVDR audit.
Conformity assessment route for high risk AI in MDR or IVDR devices
What the AI Act requires for high risk systems
Articles 8 to 15 of the AI Act set out the requirements for high risk AI systems. These are binding obligations, not guidance. Article 9 requires a risk management system that identifies and mitigates risks throughout the AI system's lifecycle, including risks to health, safety and fundamental rights. This sits alongside the ISO 14971 risk management process already required under MDR, but it is broader in scope. The AI Act risk management must address data quality, algorithmic bias, transparency failures and human oversight deficiencies. Document how these risks are assessed, mitigated and monitored.
Article 10 addresses data and data governance. Training, validation and testing datasets must be relevant, representative, free of errors and complete. Where datasets exhibit bias, that bias must be identified and addressed. For medical devices, this means documenting the demographic and clinical characteristics of training data, the methods used to detect bias, and the steps taken to ensure the AI system performs safely across the intended user population. This is not a checkbox exercise. The notified body will examine the data governance process in detail.
Article 11 requires technical documentation as specified in Annex IV. Article 12 mandates automatic recording of events, or logging, to enable traceability and post market monitoring. Article 13 requires transparency and the provision of information to deployers, meaning the instructions for use must explain how the AI system works, its limitations, and the level of accuracy. Article 14 requires human oversight measures, so that deployers can intervene, override or stop the AI system. Article 15 addresses accuracy, robustness and cybersecurity, requiring that the AI system achieves an appropriate level of performance and resilience.
These requirements are cumulative. All of them apply when the Article 6(1) test is met. They are assessed by the notified body as part of the MDR or IVDR conformity procedure. There is no separate AI Act certificate. Compliance with both regimes is reflected in the single declaration of conformity and CE marking.
Post market obligations under both regimes
Article 72 of the AI Act requires providers of high risk AI systems to establish a post market monitoring system. For MDR and IVDR products, this obligation is met by the existing post market surveillance system required under those regulations, documented in Annex III of MDR and IVDR. However, the AI Act requires that the post market monitoring system actively collects and reviews data on the AI system's performance in real world use, including any degradation in accuracy, bias that emerges in deployment, or cybersecurity incidents. This data must inform updates to the risk management system and the technical documentation.
Article 73 of the AI Act requires reporting of serious incidents. A serious incident is any incident or malfunction that directly or indirectly leads to death, serious injury or serious damage to health, property or the environment. This aligns closely with the serious incident reporting obligations under MDR Article 87 and IVDR Article 82. Where an incident involves both the medical device and the AI system, it must be reported under both regimes. In practice, the same report will cover both, but it must address the AI Act's specific requirements, including whether the incident resulted from data quality issues, algorithmic bias or transparency failures.
Post market clinical follow-up (PMCF) plans under MDR and post market performance follow-up (PMPF) plans under IVDR must now incorporate AI system performance monitoring. This includes tracking the AI system's accuracy across different patient populations, identifying any drift in model performance, and capturing user feedback on the AI system's transparency and usability. These elements feed into the periodic safety update reports (PSUR) required under MDR and IVDR. The post-market systems architecture now carries a heavier burden when AI is in scope.
What is still being settled
Several questions remain open as the AI Act and MDR/IVDR regimes begin to interact. The first is how notified bodies will interpret the data governance requirements in Article 10, particularly the obligation to ensure datasets are representative. For global medical device manufacturers, this raises the question of whether a training dataset developed in one region is sufficient for CE marking, or whether European clinical data must be included. Guidance from the Medical Device Coordination Group (MDCG) or the European Commission will clarify this, but it has not yet been published.
The second is how transparency obligations under Article 13 interact with the instructions for use requirements under MDR and IVDR. The AI Act requires that deployers receive information on the AI system's logic, limitations and expected performance. For medical devices, the instructions for use already serve this function, but the level of detail required under the AI Act may be greater. It is not yet clear whether a summary explanation suffices or whether detailed algorithmic information must be provided. Manufacturers should assume that transparency expectations will increase and prepare documentation accordingly.
The third is how the AI Act's cybersecurity requirements in Article 15 align with the cybersecurity obligations under MDR, IVDR and the EU Cybersecurity Act. The AI Act requires that high risk AI systems be resilient against attempts to manipulate them, including adversarial attacks on machine learning models. This is a specific obligation that goes beyond traditional cybersecurity threats. How notified bodies will assess this, and what evidence will satisfy them, is not yet settled. Early guidance suggests that threat modelling for AI systems, including adversarial robustness testing, will be expected.
EU AI Act application timeline for MDR and IVDR products
- 1 August 2024AI Act enters into forceRegulation (EU) 2024/1689 published and in force
- 2 February 2025Chapters I and II applyAI literacy and general provisions, including Article 4, now binding
- 2 August 2027Article 6(1) obligations applyHigh risk AI in MDR and IVDR products must meet Articles 8 to 15 requirements
- 2 August 2026Main AI Act provisions applyGeneral application date for most AI Act obligations, but Article 6(1) delayed to 2027
What this means for your team
If you manufacture a medical device or IVD that includes AI, start by applying the Article 6(1) test. Document whether the AI is the device or a safety component, and whether the device requires notified body assessment. If both conditions are met, the AI Act's high risk obligations apply from 2 August 2027. That means your technical documentation, quality management system and post market surveillance processes must be updated to address Articles 8 to 15 and Annex IV of the AI Act. This work should begin now, not in 2027. The notified body will assess AI Act compliance as part of your next MDR or IVDR audit, and they will expect evidence that the system is already in place.
For devices already on the market, consider whether a significant modification is triggered by the need to update documentation and processes to meet AI Act requirements. If the AI system itself is not changing, it is unlikely that a new conformity assessment is required, but this depends on the scope of the documentation changes. Consult your notified body early. For devices in development, integrate AI Act requirements into the design and development process from the start. Retrofitting data governance, transparency and human oversight measures after the fact is costly and time-consuming.
Regulatory affairs teams should expect that notified bodies will require training on the AI Act and may take longer to complete assessments as they develop their own internal guidance. Plan for extended review timelines in 2027 and 2028. Quality teams should audit existing processes against Articles 9, 10, 12, 14 and 15, identifying gaps and assigning owners for each workstream. Post market surveillance teams should revise PMCF and PMPF plans to capture AI system performance data, particularly accuracy, bias and cybersecurity incidents. This is not optional. It is a binding obligation under Article 72.
Key takeaways
- AI in a medical device or IVD is high risk under the EU AI Act only where the AI is the device or a safety component of it, and the device requires notified body conformity assessment. Both conditions must be met.
- Article 43(3) creates a single conformity assessment route. The notified body responsible for MDR or IVDR conformity also evaluates AI Act compliance. There is no separate AI Act assessment or certificate.
- Technical documentation must meet both MDR or IVDR Annex II requirements and AI Act Annex IV. This includes risk management, data governance, transparency, human oversight, accuracy, robustness and cybersecurity.
- Post market surveillance systems must now monitor AI system performance, including accuracy, bias and cybersecurity incidents. This feeds into PMCF, PMPF and PSUR obligations under MDR and IVDR.
- High risk AI obligations for MDR and IVDR products apply from 2 August 2027. Manufacturers should update documentation, quality systems and post market processes now, not in 2027.
Bringing AI into your own regulatory work
Understanding how the AI Act applies to your products is one thing. Deciding whether to use AI in your own regulatory operations is another. Teams building literature review tools, submission assistants or post market signal detection systems face the same questions about risk, validation, transparency and accountability. The difference is that the AI is not part of the product, so the Article 6(1) test does not apply. But the obligations around data quality, explainability and human oversight remain just as relevant.
SMEDTEC's AI as a Service is built for regulated businesses that need AI tools and agents designed around their own processes. Every implementation is validated for its intended use, reviewed by a qualified person, assessed against the EU AI Act and UK/EU GDPR, and runs on the company's own data. That means you get the efficiency gains without the compliance gaps. Whether it is automating document tagging, surfacing regulatory intelligence or structuring clinical evidence, the work is done with the same discipline you apply to the products you bring to market.
Frequently asked questions
Does AI in a medical device automatically fall under the EU AI Act?
No. Under Article 6(1) of the EU AI Act, an AI system in an MDR or IVDR product is high risk only where the AI is the product itself or a safety component of it, and that product must undergo third party conformity assessment by a notified body. If your device uses AI but does not meet both conditions, the AI Act high risk obligations do not apply.
How do MDR and the EU AI Act conformity assessments fit together?
Article 43(3) of the AI Act creates a single conformity assessment route. For high risk AI systems covered by MDR or IVDR, the provider follows the conformity assessment procedure of that legislation, and the AI Act's requirements for high risk systems (Articles 8 to 15) are assessed as part of that process. There is no separate AI Act assessment.
When do the EU AI Act obligations apply to medical device manufacturers?
The AI Act entered into force on 1 August 2024. High risk obligations for AI in MDR and IVDR products apply from 2 August 2027 under Article 113, though these dates may be amended by later legislation. Chapters I and II, including AI literacy under Article 4, apply from 2 February 2025. Check current timelines before planning.
What technical documentation is required when both MDR and the EU AI Act apply?
When an AI system in a medical device is high risk under Article 6(1), the technical documentation must meet both MDR Annex II requirements and the AI Act's Annex IV. The AI Act documentation covers risk management, data governance, transparency, human oversight, accuracy, robustness and cybersecurity. These elements are integrated into the MDR conformity assessment, not filed separately.